Page Content
"
The YSI TruLab line includes the 1110 (single channel), 1310 and 1310P (single channel) and 1320 and 1320P (dual channel) instruments providing easy-to-use and calibrate instruments ideal for the laboratory. Measure pH, mV (ORP) and temperature accurately in the laboratory with the YSI pH 1110. This benchtop instrument for basic, routine measurements in the lab is an accurate and easy-to-use solution for repeatable pH or mV measurements. Features of the 1110 include:
- Single channel measurement of pH or mV (ORP) and temperature
- AutoRead function improves repeatability
- Easy-to-read large, segmented display for pH, mV (ORP) and temperature
- Easy calibration with adjustable calibration timer
- Automatic or manual temperature compensation
- 1, 2 or 3-point calibration
- BNC connection for pH, mV; banana connection for temperature probes
- 3-year warranty
- Wide selection of electrodes - pH, ORP and temperature probes
The instrument's AutoRead function indicates when a stable measurement is reached improving repeatability.
Convenient/ReliableAn adjustable timer indicates when to perform the next calibration routine, improving measurement accuracy.
Bench: | Yes |
---|---|
Buffer Recognition: | Yes |
Certifications: | CE, electrical protective class III, cETLus, IP43 |
Display: | Segmented |
Equipment used with: | TruLab analog pH/mV/temp probes |
GLP Compliance: | Yes |
Measurement Method / Channels: | 1 Channel |
Measurement Range: | Ranges Resolutions Accuracy (+1 digit) pH -2.0 to 20.0; -2.00 to 20.00; -2.000 to 19.999 0.1; 0.01; 0.001 ±0.1; ±0.01; ±0.005 (temp of test sample 15 to 35°C; within a range of ±2 pH of the calibration point) mV -1200 to 1200.0; -2000 to 2000 0.1; 1.0 ±0.3; ±1.0 (temp of test sample 15 to 35°C) Temperature -5 to 105°C (23.0 to 221°F) 0.1 ±0.1 |
Operating pH Range: | -2.000 to 19.999, -2.00 to 20.00, -2.0 to 20.0 |
Operating Temperature: | -5 to 105°C, 23.0 to 221°F |
Parameters Measured: | pH, ORP (mV), Temperature |
Power: | universal power supply, compatible with 4 NiMH AA rechargeables. Instrument does not charge rechargeable batteries., 4 alkaline AA batteries (~2500 hours) |
Sampling: | Yes |
Storage Temperature: | -13 to 149°F, -25 to 65°C |
User Calibratable: | Yes |
User Replaceable: | Yes |
Warranty: | 3 Years |